Production of new thermoluminescent mini-dosimeters

Abstract
A method of producing CaSO4 :Dy thermoluminescent mini-dosimeters was reported in 1986 by B W Wessels for determination of the in vivo absorbed dose in radioimmunotherapy, a field in which absorbed dose gradients are important. These dosimeters, which undergo dissolution when used in a liquid environment, showed a sensitivity loss of up to 30% after 4 days of immersion in our tests. Moreover, several studies have shown that biocompatibility problems can occur during in vivo studies in animals. This paper describes the production and testing of a new type of thermoluminescent mini-dosimeter obtained by microextrusion of a mixture of LiF:Mg,Cu,P polypropylene and plastic adjuvants. These dosimeters, in the form of long 400 µm diameter filaments, can be cut to the desired length. The production process allows an LiF:Mg,Cu,P load of up to 50%. Results obtained in external irradiation indicate that these new miniature LiF:Mg,Cu,P dosimeters have good sensitivity (about 1.6 times that of CaSO4 :Dy mini-TLDs), homogeneous response within a production batch (mean ±4%), response stability in water (0.7% of variation in sensitivity after 2 weeks of immersion) and stability in aqueous solutions at different pH. LiF:Mg,Cu,P mini-dosimeters appear to be highly promising for internal dosimetry, and evaluation is in progress in animals.

This publication has 19 references indexed in Scilit: