Immunoassay for transferrin using LKB 8600 Reaction Rate Analyzer
- 1 January 1981
- journal article
- research article
- Published by Taylor & Francis in Scandinavian Journal of Clinical and Laboratory Investigation
- Vol. 41 (5) , 517-519
- https://doi.org/10.3109/00365518109090492
Abstract
A rapid immunochemical method for transferrin determination is described. The method is based on measuring turbidity caused by antigen-antibody complexes with an enzyme analyser using diluted antiserum as a starting reagent. This immunoturbidimetric method correlates well with TIBC (r=0.979; n=36). A comparison with the radial immunodiffusion method gave the correlation coefficient 0.984 and the regression line y=1. 20x-0.54 (n=36).Keywords
This publication has 8 references indexed in Scilit:
- Kinetic immunoturbidimetry: the estimation of albuminClinica Chimica Acta; International Journal of Clinical Chemistry, 1979
- A turbidimetric immuno assay (TIA) with automated individual blank compensationClinica Chimica Acta; International Journal of Clinical Chemistry, 1977
- Determination of iron and unsaturated iron-binding capacity in serum with ferrozineClinica Chimica Acta; International Journal of Clinical Chemistry, 1975
- Measurement of Serum Iron and Total Iron-Binding Capacity: Manual and Automated TechniquesPublished by Elsevier ,1972
- Determination of antisera titres using the single radial immunodiffusion methodImmunochemistry, 1969
- Immunochemical quantitation of antigens by single radial immunodiffusionImmunochemistry, 1965