Cardiac troponin T immunoassay on biotin-streptavidin-coated microplates: preliminary performance in acute myocardial infarction
- 1 January 1995
- journal article
- research article
- Published by Taylor & Francis in Scandinavian Journal of Clinical and Laboratory Investigation
- Vol. 55 (8) , 701-713
- https://doi.org/10.3109/00365519509075400
Abstract
We have adapted the cardiac troponin T (TnT) immunoassay system (ELISA troponin T, Boehringer Mannheim, Germany), which is based upon streptavidinbiotin immunoassay technology, to a sensitive microplate system. A coating of microplates with biotinylated bovine serum albumin (biotin-LC-BSA) remained stable for months. A secondary streptavidin coating was prepared as the first step of the assay. By using o-phenylenediamine (o-PD) as a substrate for the peroxidase-anti-TnT conjugate, the system allowed rapid kinetic measurement of TnT levels. The upper limit of a reference population (97.5th percentile) was found to be 0.04 μg 1−1. Intra-assay imprecision at 0.08 μg 1−1 was 8%, and 3–4% between 0.28 and 4 μg 1−1. Between-assay imprecision was 6.2% at 0.28 μg 1−1. Studies of TnT and CK-MB mass concentration in acute myocardial infarction patients, treated with streptokinase, demonstrated a clinical sensitivity of the TnT micro-plate system similar to that of the CK-MB mass concentration test, during the first 8 h after initiation of thrombolytic therapy, at discriminator levels of 0.1 μg 1−1 (TnT) and 8 μg 1−1 (CK-MB mass concentration). The early CK-MB/TnT ratio was lower in patients with signs of successful reperfusion (early peak CK-MB) than in the remaining patients (p<0.001). Serum samples from two patients with renal failure and one patient with rhabdomyolysis demonstrated strong non-linear behaviour with dilution, indicating the presence of an interfering factor. Kinetic measurement compared favourably with end-point analysis with respect to sensitivity and total analysis time. The system described greatly reduces the costs of TnT measurements compared to the ES systems. The total assay time is 70–90 min.Keywords
This publication has 20 references indexed in Scilit:
- Troponin T for the Differential Diagnosis of Ischaemic Myocardial DamageAnnals of Clinical Biochemistry: International Journal of Laboratory Medicine, 1993
- Early diagnosis of acute myocardial infarction by a newly developed rapid immunoturbidimetric assay for myoglobinHeart, 1992
- The Prognostic Value of Serum Troponin T in Unstable AnginaNew England Journal of Medicine, 1992
- New monoclonal antibodies as probes for human cardiac troponin I: Epitopic analysis with synthetic peptidesMolecular Immunology, 1992
- Early detection of acute myocardial infarction by measurement of mass concentration of creatine kinase-MBThe American Journal of Cardiology, 1991
- Diagnostic efficiency of troponin T measurements in acute myocardial infarction.Circulation, 1991
- Enzyme linked immuno assay of cardiac troponin T for the detection of acute myocardial infarction in patientsJournal of Molecular and Cellular Cardiology, 1989
- Myosin light chains release in acute myocardial infarction: non-invasive estimation of infarct sizeCardiovascular Research, 1988
- Cardiac-specific troponin-l radioimmunoassay in the diagnosis of acute myocardial infarctionAmerican Heart Journal, 1987
- Comparison of amino acid sequence of troponin I from different striated musclesNature, 1978