The use of microtiter plates for the simple and sensitive determination of insulin by an ELISA method
- 1 October 1988
- journal article
- research article
- Published by Springer Nature in Cellular and Molecular Life Sciences
- Vol. 44 (10) , 877-879
- https://doi.org/10.1007/bf01941186
Abstract
A method of insulin determination using a commercially available ELISA kit was modified for use in microtiter plates. The adapted assay, based on the binding of procine anti-guinea pig insulin antibodies to microtiter plates and insulin-peroxidase conjugate as displacer, is sensitive between 0.5 and 30 ng/ml. Since it uses only 10–40 μl of sample material it enables the determination of 5–100 pg of insulin. The rapid (5–6 h), automatable, reproducible and reliable assay makes it possible to determine many samples in a short time.This publication has 2 references indexed in Scilit: