Multimirror Scanning Spectropyrometer
- 1 March 1994
- journal article
- research article
- Published by Taylor & Francis in Journal of Modern Optics
- Vol. 41 (3) , 471-479
- https://doi.org/10.1080/09500349414550461
Abstract
A multimirror device is designed for measuring vibrational temperatures in gradient ultrasonic flows by the radiation of a heated gas in the i.r. range. The movable optical system permits one to perform spatial scanning along a jet to find temperature fields and to carry out various spectrophotometric investigations including absolute measurements of energy. The high-temperature standard of black radiation is used as a reference source. The signal of a standard source is maintained along the whole scanning site with the accuracy of ± 0·5%. The expression for determining vibrational temperatures allowing for the source brightness attenuation in the spectropyrometer illumination compartment is presented.Keywords