Signal-to-motion artifact ratio versus frequency for impedance pneumography
- 1 March 1995
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Biomedical Engineering
- Vol. 42 (3) , 321-323
- https://doi.org/10.1109/10.364521
Abstract
We measured transthoracic impedance between 12.5 and 185 kHz in nine adults. We used a system with two impedance channels, both simultaneously detecting the real part of impedance at two different frequencies. We used only two electrodes in the midaxillary line, connecting both channels in parallel. The amplitude relation between the two channels was measured for different maneuvers and frequencies. Results show for normal breathing an increase of the signal of 20% and a decrease in motion artifacts from 12.5 to 185 kHz. We conclude that, for the maneuvers studied, it is better to work at higher frequencies than the ones commonly used. Also, we suggest a method to further increase the signal-to-motion artifact ratio based on measurement at two frequencies.Keywords
This publication has 9 references indexed in Scilit:
- Dielectric properties of lung tissue as a function of air contentPhysics in Medicine & Biology, 1993
- In vivo static imaging for the real and the reactive parts in electrical impedance tomography using multifrequency techniquesPublished by Institute of Electrical and Electronics Engineers (IEEE) ,1992
- The electrode system in impedance-based ventilation measurementIEEE Transactions on Biomedical Engineering, 1992
- Two-frequency impedance plethysmograph: real and imaginary partsMedical & Biological Engineering & Computing, 1990
- Electrode Motion Artifacts in Electrical Impedance PneumographyIEEE Transactions on Biomedical Engineering, 1985
- Electrical resistivity of lung at 100 kHzMedical & Biological Engineering & Computing, 1967
- Impedance pneumography—A survey of instrumentation techniquesMedical & Biological Engineering & Computing, 1966
- Volumetric dynamics of respiration as measured by electrical impedance plethysmographyJournal of Applied Physiology, 1964