Interference by ascorbic acid in test systems involving peroxidase. I. Reversible indicators and the effects of copper, iron, and mercury.
Open Access
- 1 April 1982
- journal article
- research article
- Published by Oxford University Press (OUP) in Clinical Chemistry
- Vol. 28 (4) , 578-588
- https://doi.org/10.1093/clinchem/28.4.578
Abstract
I describe the mechanism whereby ascorbic acid can hamper test systems involving a peroxide-generating system, peroxidase, and a benzidine-type indicator. In test media such as urines, abnormally high concentrations of ascorbic acid may give rise to false negatives in the determination of analytes such as glucose. Absorbance measurements in solution or reflectance measurements on commercially available paper reagent strips demonstrate either inhibition of visible activity by ascorbic acid or a lag time in the development of oxidized indicator color. The duration of the lag time is proportional to the ascorbic acid concentration, is inversely proportional to the enzyme concentration, and is also affected by concentrations of hydrogen peroxide and o-tolidine indicator. The same results were seen in both citrate buffer pH 5 and phosphate buffer pH 7. Because the complete system (o-tolidine indicator, hydrogen peroxide, and peroxidase) must be present if the ascorbate is to be oxidized rapidly, this indicates that ascorbic acid inhibits color development by re-reducing oxidized indicator as fast as it is formed; the o-tolidine then acts catalytically in oxidizing ascorbic acid. Added Cu2+ and Fe3+, both known to react with ascorbic acid, had measurable but small effects on the system. In contrast, Hg2+ abolished the ascorbic acid-elicited lag time, even when present in near-stoichiometric concentration. Hg2+ showed little inhibitory effect on the activity of either glucose oxidase or peroxidase. Presumably it rapidly oxidizes ascorbic acid to dehydroascorbate. The reaction of mercuric ion with ascorbate was measured by reflectance measurements of paper reagent strips in addition to absorbance measurements of solution assays; equivalent results were obtained. Incorporation of Hg2+ into reagent strips can thus render both strips and solution diagnostic tests insensitive to interfering substances such as ascorbic acid.This publication has 9 references indexed in Scilit:
- Urinary Glucose and Vitamin CAmerican Journal of Clinical Pathology, 1977
- The role of ascorbic acid in the regulation of cholesterol metabolism and in the pathogenesis of atherosclerosisAtherosclerosis, 1976
- Purification and Properties of the Glucose Oxidase from Aspergillus nigerJournal of Biological Chemistry, 1965
- 3,4-DIHYDROXYPHENYLETHYLAMINE BETA-HYDROXYLASE - PHYSICAL PROPERTIES COPPER CONTENT AND ROLE OF COPPER IN CATALYTIC ACTIVITY1965
- Identification, by Electron Paramagnetic Resonance Spectroscopy, of Free Radicals Generated from Substrates by PeroxidaseJournal of Biological Chemistry, 1960
- ASCORBIC ACID IN AROMATIC HYDROXYLATION .1. A MODEL SYSTEM FOR AROMATIC HYDROXYLATION1954
- Oxidation Processes. XVII.1 The Autoxidation of Ascorbic Acid in the Presence of CopperJournal of the American Chemical Society, 1944
- The catalytic decomposition of hydrogen peroxide by iron saltsProceedings of the Royal Society of London. Series A. Mathematical and Physical Sciences, 1934
- The kinetics of peroxidase actionBiochemical Journal, 1931