Performance characteristics of a Kodak computed radiography system
- 5 January 1999
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 26 (1) , 27-37
- https://doi.org/10.1118/1.598781
Abstract
The performance characteristics of a photostimulable phosphor based computed radiographic (CR) system were studied. The modulation transfer function (MTF), noise power spectra (NPS), and detective quantum efficiency (DQE) of the Kodak Digital Science computed radiography (CR) system (Eastman Kodak Co.-model 400) were measured and compared to previously published results of a Fuji based CR system (Philips Medical Systems-PCR model 7000). To maximize comparability, the same measurement techniques and analysis methods were used. The DQE at four exposure levels (30, 3, 0.3, 0.03 mR) and two plate types (standard and high resolution) were calculated from the NPS and MTF measurements. The NPS was determined from two-dimensional Fourier analysis of uniformly exposed plates. The presampling MTF was determined from the Fourier transform (FT) of the system's finely sampled line spread function (LSF) as produced by a narrow slit. A comparison of the slit type ("beveled edge" versus "straight edge") and its effect on the resulting MTF measurements was also performed. The results show that both systems are comparable in resolution performance. The noise power studies indicated a higher level of noise for the Kodak images (approximately 20% at the low exposure levels and 40%-70% at higher exposure levels). Within the clinically relevant exposure range (0.3-3 mR), the resulting DQE for the Kodak plates ranged between 20%-50% lower than for the corresponding Fuji plates. Measurements of the presampling MTF with the two slit types have shown that a correction factor can be applied to compensate for transmission through the relief edges.Keywords
This publication has 9 references indexed in Scilit:
- Physical measures of image quality in photostimulable phosphor radiographic systemsPublished by SPIE-Intl Soc Optical Eng ,1997
- Clinical experiences with computed radiographyEuropean Journal of Radiology, 1996
- DQE(f) of four generations of computed radiography acquisition devicesMedical Physics, 1995
- Computed radiography in pediatricsJournal of Digital Imaging, 1995
- Effects of undersampling on the proper interpretation of modulation transfer function, noise power spectra, and noise equivalent quanta of digital imaging systemsMedical Physics, 1995
- Bedside chest radiography.Radiology, 1994
- A simple method for determining the modulation transfer function in digital radiographyIEEE Transactions on Medical Imaging, 1992
- Quantitative radiographic imaging using a photostimulable phosphor systemMedical Physics, 1990
- Imaging performance of a digital storage phosphor systemMedical Physics, 1987