A comparison of double‐tuned surface coils
- 31 May 1989
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 10 (3) , 302-309
- https://doi.org/10.1002/mrm.1910100303
Abstract
The combined acquisition of proton images and localized spectra is considered essential to the practical application of NMR techniques to human and animal research. Double‐tuned surface coils which have been introduced to the literature are intended to address the problem: however, a careful evaluation of available designs is lacking. The “trap” method, the loop gap resonator design, and the transformer‐coupled double‐tuned design are evaluated here using bench tests of signal intensity and Q as well as signal‐to‐ noise measurements on a 2‐T imager/spectrometer. Comparisons are made relative to optimized single‐tuned circuits of the same size for both protons at 85 MHz and phosphorus at 34 MHz. The results suggest that the “trap” design and the transformer coupled design are very efficient (98%) in the low‐frequency mode (34 MHz) while the loop gap resonator is relatively inefficient (82%). In the high‐frequency mode (85 MHz) the loop gap resonator is 75% efficient while the “trap” design and the transformer coupled coil are closer to 50% efficient. © 1989 Academic press. Inc. © 1989 Academic Press, Inc.This publication has 9 references indexed in Scilit:
- A transformer‐coupled double‐resonant probe for NMR imaging and spectroscopyMagnetic Resonance in Medicine, 1987
- The design and use of a dual-frequency surface coil providing proton images for improved localization in 3 1 P spectroscopy of small lesionsMedical Physics, 1986
- Transmission line matching of surface coilsMagnetic Resonance in Medicine, 1986
- Design and evaluation of a radio frequency coil for nuclear magnetic resonance imaging of fluorine and protonsMedical Physics, 1985
- A new double-tuned probed for concurrent 1H and 31P NMRJournal of Magnetic Resonance (1969), 1985
- A technique for simultaneous 1H and 31P NMR at 2.2 T in VivoJournal of Magnetic Resonance (1969), 1985
- High-field 13C NMR spectroscopy of tissue in Vivo. A double-resonance surface-coil probeJournal of Magnetic Resonance (1969), 1984
- An improved tune and match circuit for B0 shimming in intact biological samplesJournal of Magnetic Resonance (1969), 1982
- Single coil probe with transmission-line tuning for nuclear magnetic double resonanceReview of Scientific Instruments, 1976