Safe Performance of Magnetic Resonance Imaging on Five Patients with Permanent Cardiac Pacemakers
- 30 June 1996
- journal article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 19 (6) , 913-919
- https://doi.org/10.1111/j.1540-8159.1996.tb03387.x
Abstract
Five patients with permanent cardiac pacemakers (Pacesetter models 261, 285, 2016, 2020, 2022) underwent magnetic resonance imaging (MRI). Only one patient (underlying rhythm asystole) was pacemaker dependent. A variety of pacing configurations (single and dual chamber; unipolar and bipolar; sensor and nonsensor driven) were scanned. A thorough evaluation of each pacing system was performed before and after scanning including determination of pacing and sensing thresholds. During MRI the patient was monitored using either ECG, pulse oximetry, or direct voice contact. In four patients heavy dressings were applied over the pacemaker pocket site. Patients were asked to report any symptoms experienced during MRI. The four nonpacemaker dependent patients remained in sinus rhythm throughout the MRI. During and after the MRI all pacemakers continued to function normally except for one transient pause of approximately 2 seconds (noted by pulse oximeter) toward the end of the scan. This occurred in a pacemaker dependent patient with a unipolar dual chamber device programmed DOO. No patient experienced any torque or heat sensation. When appropriate strategies are used our experience suggests that MRI may be performed, when necessary, with an acceptable risk-benefit ratio to the patient. It is unclear whether the isolated pause that was observed was due to the effect of the MRI, an artifact with the monitoring system, or oversensing by the pacemaker. Appropriate patient selection, close monitoring during the scan, and follow-up after MRI are of paramount importance. Further study is necessary to refine the appropriate strategies that could be used to consistently perform MRI safely in a selected pacemaker population.Keywords
This publication has 12 references indexed in Scilit:
- The Clinical Efficacy of Magnetic Resonance Imaging in NeuroimagingAnnals of Internal Medicine, 1994
- The usefulness of cardiovascular magnetic resonance imagingCurrent Problems in Cardiology, 1994
- Case Report: Nuclear Magnetic Resonance Imaging in a Patient with a PacemakerThe Lancet Healthy Longevity, 1993
- Safety of Patients with Medical Devices during Application of Magnetic Resonance MethodsAnnals of the New York Academy of Sciences, 1992
- MR imaging and metallic implants for anterior cruciate ligament reconstruction: Assessment of ferromagnetism and artifactJournal of Magnetic Resonance Imaging, 1992
- Nuclear Magnetic Resonance Imaging in a Patient with a DDD PacemakerPacing and Clinical Electrophysiology, 1989
- Effect of 1.5 tesla nuclear magnetic resonance imaging scanner on implanted permanent pacemakersJournal of the American College of Cardiology, 1987
- Effect of magnetic resonance imaging on DDD pacemakersThe American Journal of Cardiology, 1986
- The Effects of Nuclear Magnetic Resonance Imagers on External and Implantable Pulse GeneratorsPacing and Clinical Electrophysiology, 1984
- The effects of nuclear magnetic resonance on patients with cardiac pacemakers.Radiology, 1983