The oxygen status of the arterial blood revised: Relevant oxygen parameters for monitoring the arterial oxygen availability
- 1 January 1990
- journal article
- case report
- Published by Taylor & Francis in Scandinavian Journal of Clinical and Laboratory Investigation
- Vol. 50 (sup203) , 17-28
- https://doi.org/10.3109/00365519009087488
Abstract
The new generation of very accurate multi-wavelength oximeters, e.g. OSM3, for in vitro measurement of the hemoglobin oxygen saturation, total hemoglobin concentration, and carboxy- and methemoglobin fractions opens new aspects of oxygen monitoring. Combined with the data from the blood gas analyzer (e.g. ABL300) these very accurate measurements allow the calculation of several derived oxygen parameters on the basis of a set of newly developed calculation algorithms. The traditional parameters obtained from an arterial sample are the oxygen tension (pO2) and the hemoglobin oxygen saturation (sO2). Clinical examples illustrate that the pO2 and the sO2 even in combination may give misleading information. The new algorithm calculates three extra oxygen parameters. 1) The oxygen extraction tension, px, defined as the tension required to extract 2.3 mmol of oxygen per liter blood. It signals the mixed venous pO2 level on the assumption that the arterio-venous oxygen difference is normal (2.3 mmol/L). 2) The concentration of extractable oxygen, cx, defined as the concentration of oxygen extracted at a tension of 5.0 kPa. 3) The oxygen compensation factor, Qx, derived as (2.3 mmol/L)/cx. It may be interpreted as the increase in cardiac output necessary to maintain a normal mixed venous pO2 of 5 kPa. These three parameters indicate the oxygen availability of the blood and summarize important properties of the arterial blood in relation to oxygen supply of the tissues, including the arterial pO2, the 'active' hemoglobin concentration (equivalent to the oxygen capacity), and the hemoglobin oxygen affinity (p50). The set of data measured with the blood gas analyzer, e.g. the ABL300 combined with the data measured with the OSM3 contains much more information than is routinely utilized. This information is extracted and summarized by our calculation algorithm. Omitting the calculation of the extra oxygen parameters involves a risk of losing valuable information.Keywords
This publication has 9 references indexed in Scilit:
- The oxygen status algorithm: A computer program for calculating and displaying pH and blood gas dataScandinavian Journal of Clinical and Laboratory Investigation, 1990
- Arterial oxygen status determined with routine pH/blood gas equipment and multi-wavelength hemoximetry: Reference values, precision, and accuracyScandinavian Journal of Clinical and Laboratory Investigation, 1990
- Cellular Oxygen Metabolism During Sepsis and ShockPublished by American Medical Association (AMA) ,1988
- Measured and derived quantities with modern pH and blood gas equipment: Calculation algorithms with 54 equationsScandinavian Journal of Clinical and Laboratory Investigation, 1988
- ADDITIONAL BLOOD-GAS VARIABLES FOR THE RATIONAL CONTROL OF OXYGEN THERAPY †British Journal of Anaesthesia, 1987
- Circulatory mechanisms of shock and their mediatorsCritical Care Medicine, 1987
- Spectrophotometric determination of hemoglobin pigments in neonatal bloodClinica Chimica Acta; International Journal of Clinical Chemistry, 1987
- Diode-array spectrophotometry for simultaneous measurement of hemoglobin pigmentsClinica Chimica Acta; International Journal of Clinical Chemistry, 1987
- The concept of a critical oxygen deliveryIntensive Care Medicine, 1987