Intrinsic and routine quality of serum total potassium measurement as investigated by split-sample measurement with an ion chromatography candidate reference method
- 1 April 1998
- journal article
- research article
- Published by Oxford University Press (OUP) in Clinical Chemistry
- Vol. 44 (4) , 849-857
- https://doi.org/10.1093/clinchem/44.4.849
Abstract
We evaluated the intrinsic quality of eight routine test systems for the measurement of serum total potassium (K+), as well as the routine quality of four of these systems, using a group of 60 single-donation serum samples that had been certified with an ion chromatography reference method. The intrinsic quality of the tests was evaluated by analysis of the sera in the manufacturers’ application laboratories under strict internal quality control. The routine quality was evaluated by analysis of the same sera in five (per system) routine laboratories under daily working conditions. The results of the study were interpreted in light of the most stringent specifications derived from the biological variation of K+, which require limits of 6.3% for total error and 1.6% for systematic error. The study revealed that the intrinsic quality of all systems was excellent. None of the test systems yielded a substantial number of results outside the 6.3% total error limit, and only one test system exceeded the 1.6% systematic error limit. The majority of the routine laboratories reproduced the manufacturers’ intrinsic quality. In particular, most laboratories satisfied the 6.3% total error limit. However, several laboratories exceeded the 1.6% systematic error limit. Generally, there was a considerable difference in quality between the participating laboratories. This showed that the major problems for serum K+ analysis (for samples with no unusual matrices and with concentrations within the reference interval) are at the routine laboratory.Keywords
This publication has 8 references indexed in Scilit:
- Graphical interpretation of analytical data from comparison of a field method with a Reference Method by use of difference plotsClinical Chemistry, 1997
- Practical aspects of suppressed ion chromatography for cations in human serumJournal of Chromatography A, 1997
- Validation of candidate reference methods based on ion chromatography for determination of total sodium, potassium, calcium and magnesium in serum through comparison with flame atomic emission and absorption spectrometryClinical Biochemistry, 1996
- Ion chromatography as potential reference methodology for the determination of total sodium and potassium in human serumJournal of Chromatography A, 1995
- Development of criteria for the evaluation of reference method valuesScandinavian Journal of Clinical and Laboratory Investigation, 1993
- A model for a comprehensive measurement system in clinical chemistry.Clinical Chemistry, 1979
- Rationale for Reference Methods in Clinical ChemistryPublished by Walter de Gruyter GmbH ,1976
- An Idea Whose Time Has ComeClinical Chemistry, 1973