Pre-analytical variation of some haematological quantities
- 1 January 2008
- journal article
- Published by Walter de Gruyter GmbH in cclm
- Vol. 46 (8) , 1168-1170
- https://doi.org/10.1515/cclm.2008.229
Abstract
It is customary to assume that pre-analytical variation is negligible if all pre-analytical sources of variation have been standardised. The aim of this study was to quantify the pre-analytical variation of some haematological quantities frequently measured in clinical laboratories.The experimental design considers different sources of pre-analytical variation that usually occur day-to-day for samples from patients.The results demonstrate that for concentrations of erythrocytes, haemoglobin, leukocytes, neutrophilocytes (segmented) and lymphocytes, and for the volume fraction of erythrocytes ("packed cell volume") in blood, the pre-analytical variance is not negligible.We suggest that these variances should be taken into account when estimating the uncertainty of measurement for patient results.Keywords
This publication has 3 references indexed in Scilit:
- Description of examinations and their results and ISO standard 15189cclm, 2006
- Evaluation of Uncertainty of Measurement in Routine Clinical Chemistry - Applications to Determination of the Substance Concentration of Calcium and Glucose in Serumcclm, 2002
- Generation and Application of Data on Biological Variation in Clinical ChemistryCritical Reviews in Clinical Laboratory Sciences, 1989