The tracheal tube with a high-volume, low-pressure cuff at various airway inflation pressures
- 1 November 1998
- journal article
- clinical trial
- Published by Wolters Kluwer Health in European Journal of Anaesthesiology
- Vol. 15 (6) , 629-632
- https://doi.org/10.1097/00003643-199811000-00002
Abstract
When the tracheal tube with a high-volume, low-pressure cuff is used, the pressure exerted by the cuff on the tracheal wall is similar to the intracuff pressure, and the pressure is claimed to be lower than the capillary perfusion pressure; however, it is not known if this is the case when a high airway pressure is required. In a randomized, cross-over design, we studied 61 patients (31 men) and measured the intracuff pressures of the tracheal tube at various airway pressures. While the lungs were inflated at five different airway pressures in turn (10, 15, 20, 25 and 30 cm H2O), the cuff of the tracheal tube was inflated with air until it prevented an airleak ('just-seal' pressure). Intracuff pressures [median (range)] at airway pressures of 10, 15, 20, 25 and 30 cm H2O were 8 (0-20), 15 (4-20), 22 (6-32), 26 (11-52) and 31 (16-54) cm H2O, respectively, for men and 6 (0-20), 11 (0-20), 15 (0-24), 21 (0-32) and 25 (1-41) cm H2O, respectively, for women. Therefore, we conclude that the pressure exerted by the cuff of the tracheal tube on the tracheal wall is unlikely to exceed the capillary perfusion pressure (arbitrarily defined as 25 mmHg or 34 cm H2O) when the airway pressure is 25 cm H2O or less, but it may exceed the capillary perfusion pressure when the airway pressure is greater than 25 cm H2O.Keywords
This publication has 10 references indexed in Scilit:
- Influence of airway pressure on minimum occlusive endotracheal tube cuff pressureCritical Care Medicine, 1997
- High-volume, Low-pressure CuffsChest, 1991
- Tracheal tube cuff pressureAnaesthesia, 1988
- RATIONALIZATION OF THE SELECTION OF TRACHEAL TUBESBritish Journal of Anaesthesia, 1986
- Intracuff Pressures in Endotracheal and Tracheostomy TubesChest, 1985
- Endotracheal cuff pressure and tracheal mucosal blood flow: endoscopic study of effects of four large volume cuffs.BMJ, 1984
- Tracheal size and shape: effects of change in intraluminal pressure.Radiology, 1983
- Prevention of complications from prolonged tracheal intubationThe American Journal of Surgery, 1978
- Performance Characteristics of Tracheal CuffsInternational Anesthesiology Clinics, 1974
- Intratracheal CuffsAnesthesiology, 1969