In Vivo Assessment of Stent Expansion and Recoil in Normal Porcine Coronary Arteries
- 17 August 1999
- journal article
- research article
- Published by Wolters Kluwer Health in Circulation
- Vol. 100 (7) , 756-760
- https://doi.org/10.1161/01.cir.100.7.756
Abstract
Background—Despite the routine use of high pressure, coronary stents generally fail to achieve a cross-sectional area (CSA) >60% to 80% of the nominal CSA of their dilating balloon. The extent to which incomplete balloon expansion and postdeflation stent recoil contribute to this failure has not been fully evaluated. Methods and Results—Thirty-two stents (8 Gianturco-Roubin II, 8 Palmaz-Schatz, 8 MultiLink, and 8 NIR) were deployed in nondiseased coronary arteries of 8 Yorkshire pigs. All stents were then expanded according to 1 of 3 balloon strategies: appropriately sized compliant balloons, oversized compliant balloons, or oversized noncompliant balloons. Continuous ultrasound imaging was performed during stepwise balloon inflation and deflation, with an 0.018-in imaging core positioned within the guidewire lumen of the balloon. In these normal arteries, balloon underexpansion relative to the nominal size was not observed. After balloon deflation, however, all stents showed significant recoil from their maximum inflated CSA. Recoil was significantly greater for the coil Gianturco-Roubin II stent (30% CSA) than for the 3 slotted-tube stent designs (15% to 17% CSA). Conclusions—In normal, compliant coronary arteries, stent recoil is the predominant mechanism by which stents fail to achieve the nominal CSA of their dilating balloon. The magnitude of this in vivo stent recoil is significantly greater than reported from bench testing and varies with stent design (coil versus slotted tube). Postdilatation strategies that result in controlled stent overexpansion are needed to overcome this recoil phenomenon and result in a final stent CSA that approximates the reference artery CSA.Keywords
This publication has 14 references indexed in Scilit:
- Clinical, resource and cost benefits of the RESTORE study applied to US populationJournal of the American College of Cardiology, 1998
- Intravascular ultrasound-guided optimized stent deployment Immediate and 6 months clinical and angiographic results from the Multicenter Ultrasound Stenting in Coronaries Study (MUSIC Study)European Heart Journal, 1998
- Mechanisms of Residual Lumen Stenosis After High-Pressure Stent ImplantationCirculation, 1998
- Subacute Stent Thrombosis in the Era of Intravascular Ultrasound-Guided Coronary Stenting Without Anticoagulation: Frequency, Predictors and Clinical OutcomePublished by Elsevier ,1998
- Acute and Long-Term Outcome After Palmaz-Schatz Stenting: Analysis From the New Approaches To Coronary Intervention (NACI) RegistryThe American Journal of Cardiology, 1997
- Small stent size and intimal hyperplasia contribute to restenosis: A volumetric intravascular ultrasound analysisJournal of the American College of Cardiology, 1995
- Intracoronary Stenting Without Anticoagulation Accomplished With Intravascular Ultrasound GuidanceCirculation, 1995
- Intracoronary ultrasound observations during stent implantation.Circulation, 1994
- Quantitative analysis of elastic recoil after balloon angioplasty and after intracoronary implantation of balloon-expandable Palmaz-Schatz stentsJournal of the American College of Cardiology, 1993
- The importance of acute luminal diameter in determining restenosis after coronary atherectomy or stenting.Circulation, 1992