Defects in failed stems of hip prostheses
- 1 September 1978
- journal article
- research article
- Published by Wiley in Journal of Biomedical Materials Research
- Vol. 12 (5) , 635-651
- https://doi.org/10.1002/jbm.820120506
Abstract
A study has been made of the fractures and associated material from 34 prostheses broken in service. All the fractures appear to be of the fatigue type. Search by me-tallographic analysis for cracklike generating defects in material near the fracture reveals such conditions as abnormally coarse crystals, large nonmetallic inclusions, large inclusion population, undissolved master alloy particles, abnormal carbide segregation, interdendritic segregation, shrinkage, and gas porosity. Recommendations are made for the choice of metallic alloys for femoral stem application.This publication has 10 references indexed in Scilit:
- Nondestructive/Destructive Test Correlations and Fracture Mechanics AnalysisJournal of Testing and Evaluation, 1977
- A comparative experimental study of stresses in femoral total hip replacement components: The effects of prostheses orientation and acrylic fixationJournal of Biomechanics, 1976
- A statistical treatment of fatigue of the cast Co‐Cr‐Mo prosthesis alloyJournal of Biomedical Materials Research, 1976
- Fracture of Femoral Prostheses in Total Hip ReplacementClinical Orthopaedics and Related Research, 1975
- Fatigue fractures of the femoral component of Charnley and Charnley‐Müller type total hip prosthesesJournal of Biomedical Materials Research, 1975
- Factors in the Mechanical Failure of the Femoral Component in Total Hip Prosthesis:Report of Six Fatigue Fractures of the Femoral Stern and Results of Experimental Loading TestsActa Orthopaedica, 1974
- Corrosion-related Failures in Metallic ImplantsPublished by Wolters Kluwer Health ,1972
- Influence of nonequilibrium second-phase particles formed during solidification upon the mechanical behavior of an aluminum alloyMetallurgical Transactions, 1971