Abstract
A computer-assisted planning and visualization system (the Orthopaedic Surgery Planning System) was tested for pedicle screw insertion in vivo. To evaluate the system's applicability for regular intraoperative use and its accuracy for pedicle screw placement in vivo. Pedicle screw placement poses considerable anatomic and biomechanical risks. The reported rates of screw misplacement with conventional insertion techniques are unacceptably high. It previously has been shown in vitro that computer assistance offers the potential to decrease the number of screws perforating the pedicular cortex. The accuracy of 162 pedicle screws inserted with the Orthopaedic Surgery Planning System was assessed by means of postoperative computed tomography evaluation. Reconstructions of the horizontal, frontal, and sagittal planes were analyzed. Cortex perforations were graded in steps of 2 mm. The cortex was perforated in 2.7% of pedicles. Complete preoperative computed tomography scanning of the levels to be operated on is essential to allow for a precise image reconstruction. Initial difficulties in applying the system contribute to the malplacements. A learning curve for general handling of the Orthopaedic Surgery Planning System was observed. The system provides a safe and reproducible technique for pedicle screw insertion. Other applications in the field of spine surgery are under evaluation.Keywords
This publication has 35 references indexed in Scilit:
- Image-Guided Insertion of Transpedicular ScrewsSpine, 1995
- Measurement of Vertebral Cortical Integrity During Pedicle Exploration for Intrapedicular FixationSpine, 1995
- Pedicle Screw Placement With Intraosseous EndoscopySpine, 1994
- CT Analysis of Pedicles and Screw Tracts After Implant Removal in Thoracolumbar FracturesJournal of Spinal Disorders, 1993
- Location of transpedicular screws for fixation of the lower thoracic and lumbar spine: Computed tomography of 45 fracture casesActa Orthopaedica, 1993
- Placement of Transpedicular Vertebral Screws Close to Anterior Vertebral CortexSpine, 1989
- Roentgenographic Measurement of Pedicle Screw PenetrationClinical Orthopaedics and Related Research, 1989
- Early Results of Spinal Fusion Using Variable Spine Plating SystemSpine, 1988
- An Internal Fixator for Posterior Application to Short Segments of the Thoracic, Lumbar, or Lumbosacral Spine Design and TestingPublished by Wolters Kluwer Health ,1986
- Internal Fixation of the Lumbar Spine with Pedicle Screw PlatingClinical Orthopaedics and Related Research, 1986