Reprogramming of Implanted Pacemaker Following External Defibrillation
- 1 October 1978
- journal article
- case report
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 1 (4) , 514-520
- https://doi.org/10.1111/j.1540-8159.1978.tb03514.x
Abstract
This report describes reprogramming of a pulse generator consequent to cardiac defibrillation. Analysis of the explunted pulse generator revealed normal function. We tested the electrical properties of several defibrillaiors and our studies suggest that erratic electrostatic discharges at the time of paddle application before defibrillation may create a signal sequence capable of reprogrumming the particular pulse generator by activating either the reed switch or internal electronics.Keywords
This publication has 5 references indexed in Scilit:
- Phantom Pacemaker Programming*Pacing and Clinical Electrophysiology, 1978
- Preliminary Experience with the Use of a Programmable PacemakerChest, 1975
- Pacemaker Failure following External DefibrillationCirculation, 1972
- Pacemaker Failure Following External DefibrillationCirculation, 1971
- Protection of implanted pacemakers from excessive electrical energy of D.C. shockThe American Journal of Cardiology, 1969