The Margin of Safety in the Use of a Straight Path Centrifugal Blood Pump

Abstract
A new centrifugal blood pump with a rotor that arranges 6 straight paths radially was developed for open heart surgery and temporary circulatory support. We describe comparative studies of the margin of safety in the practical use of the new pump. This pump was evaluated for temperature increase, cavitation, and pressure sensitivity. Two commercially available centrifugal pumps, the Biomedicus cone type and the Sarns 3M impeller type, were used as control pumps. The temperature increase in the new pump was four times slower than in the impeller pump when the outlet and the inlet of the pump was clamped. No sign of cavitation was observed when 0.1 ml air was introduced to the pump inlet under a negative pressure of 200 mm Hg in fresh bovine blood. As for pressure sensitivity of centrifugal pumps in practical applications, circuit resistance was a more essential factor than flow-pressure curves of the pump.