The IPEMB code of practice for electron dosimetry for radiotherapy beams of initial energy from 2 to 50 MeV based on an air kerma calibration
- 1 December 1996
- journal article
- guideline
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 41 (12) , 2557-2603
- https://doi.org/10.1088/0031-9155/41/12/001
Abstract
This report contains the recommendations of the Electron Dosimetry Working Party of the UK Institution of Physics and Engineering in Medicine and Biology (IPEMB). The recommendations consist of a code of practice for electron dosimetry for radiotherapy beams of initial energy from 2 to 50 MeV. The code is based on the 2 MV (or 60Co) air kerma calibration of the NE 2561/2611 chamber, which is used as the transfer instrument between national standards laboratory and hospitals in the UK. The code utilizes an ND.air approach. Designated chambers are the NE 2571 (graphite-walled Farmer chamber), to be calibrated against the transfer instrument in a megavoltage photon beam, and three parallel-plate chambers, to be calibrated against the NE 2571 in a higher-energy electron beam. The practical code is supplemented by comprehensive discussion of the theoretical background and the sources and values of included data.Keywords
This publication has 66 references indexed in Scilit:
- Ion recombination corrections for plane-parallel and thimble chambers in electron and photon radiationPhysics in Medicine & Biology, 1993
- Addendum to the code of practice for electron beam dosimetry in radiotherapy (1985): interim additional recommendationsPhysics in Medicine & Biology, 1992
- Implication of electron backscattering for electron dosimetryPhysics in Medicine & Biology, 1991
- Influence of energy and angular spread on stopping-power ratios for electron beamsPhysics in Medicine & Biology, 1989
- Comparison of ionisation measurements in water and polystyrene for electron beam dosimetryPhysics in Medicine & Biology, 1985
- The collection efficiency of an ionization chamber in a pulsed and magnetically swept electron beam: Limits of validity of the two‐voltage techniqueMedical Physics, 1984
- Dose errors due to charge storage in electron irradiated plastic phantomsMedical Physics, 1984
- Density Effect in Liquid WaterRadiation Research, 1982
- The National Physical Laboratory secondary standard therapy-level X-ray exposure meterThe British Journal of Radiology, 1972
- Einfluß der Vielfachstreuung von Elektronen auf die Ionisation in gasgefüllten HohlräumenRadiation and Environmental Biophysics, 1968