Correction of scatter in computed tomography images of bone
- 1 May 1987
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 14 (3) , 335-340
- https://doi.org/10.1118/1.596089
Abstract
A cylindrical aluminium/Plexiglas phantom representing trabecular bone surrounded by various amounts of cortical bone was constructed. Measurements of this phantom using a computed tomography scanner with a 125I photon source demonstrated errors of 0% to 28% in the density of trabeculae bone. Two contributing factors are identified: scatter and exponential edge-gradient effect. A simple first-order correction is developed to correct for the scatter-induced error. Relative to the exponential edge-gradient effect, which contributes up to 3.4% error over the range of cortical thicknesses measured, the correction procedure reduces the scatter-induced error to a level of -0.66% to +0.61%. The consistency of the optimized correction parameters with the physical model as well as the effect of scatter measured by the same phantom on a GE 8800 scanner are shown.Keywords
This publication has 10 references indexed in Scilit:
- Effects of scattered X-rays on CT imagesPhysics in Medicine & Biology, 1985
- A Postprocessing Dual Energy Technique for Vertebral CT DensitometryJournal of Computer Assisted Tomography, 1984
- The influence of scattered radiation on the CT numbers of bone on a scanner with a fixed detector arrayPhysics in Medicine & Biology, 1983
- Quantitative Measurement of Bone Density Using Gamma-Ray Computed TomographyJournal of Computer Assisted Tomography, 1982
- Compton scatter effects in CT reconstructionsMedical Physics, 1982
- The effects of scatter in x‐ray computed tomographyMedical Physics, 1982
- Scattered radiation in fan beam imaging systemsMedical Physics, 1982
- The exponential edge-gradient effect in X-ray computed tomographyPhysics in Medicine & Biology, 1981
- Precise Measurement of Vertebral Mineral Content Using Computed TomographyJournal of Computer Assisted Tomography, 1980
- Standardization of Computed Tomography Images by Means of a Material-Selective Beam Hardening CorrectionJournal of Computer Assisted Tomography, 1978