Hybrid Mercuric Iodide (HgI2) - Gadolinium Orthosilicate (GSO) Detector for PET
- 1 January 1985
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Nuclear Science
- Vol. 32 (1) , 533-537
- https://doi.org/10.1109/TNS.1985.4336888
Abstract
Gadolinium orthosilicate (GSO), a high density (6.71 g/cm3), high effective Z (59) scintillation detector with faster decay time (60 ns) and higher light yield than bismuth germanate (BGO), was evaluated as a replacement for BGO for a hybrid detection system for high resolution Positron Emission Tomography (PET). The detection system consists of multiple GSO detectors attached to a single photomultiplier tube (PMT) and a HgI2 photodetector attached to each GSO crystal. The PMT signal provides coincidence timing and energy discrimination and the photodetector signal identifies the crystal of interaction. GSO light yield was 1.7 times that of BGO with energy resolution consistent with improved photon statistics (17.8 to 13.8% FWHM). Resolution of GSO coupled to a HgI2 photodetector was 13.6% FWHM. Coincidence timing was 2.3 ns FWHM. Timing between PMT and HgI2 was 136 ns FWHM.Keywords
This publication has 11 references indexed in Scilit:
- Initial Characterization of a Bgo-Photodiode Detector for High Resolution Positron Emission TomographyIEEE Transactions on Nuclear Science, 1984
- Large Area Mercuric Iodide PhotodetectorsIEEE Transactions on Nuclear Science, 1984
- PERFORMANCE EVALUATION OF A POSITRON TOMOGRAPH DESIGNED FOR BRAIN IMAGINGJournal of Computer Assisted Tomography, 1983
- Cerium-activated Gd2SiO5 single crystal scintillatorApplied Physics Letters, 1983
- A Positron Tomograph Employing a One Dimension BGO Scintillation CameraIEEE Transactions on Nuclear Science, 1983
- A High Resolution Detection System for Positron TomographyIEEE Transactions on Nuclear Science, 1983
- Investigation of a Technique for Providing a Pseudo-Continuous Detector Ring for Positron TomographyIEEE Transactions on Nuclear Science, 1982
- Emission Computed Tomography SimulatorIEEE Transactions on Nuclear Science, 1980
- Design and Performance Characteristics of a Positron Emission Computed Axial Tomograph--ECAT®-IIIEEE Transactions on Nuclear Science, 1979
- Improved Timing with NaI(Tl)IEEE Transactions on Nuclear Science, 1966