A Prospective Randomized Cross‐Over Comparison of Mono‐ and Biphasic Defibrillation Using Nonthoracotomy Lead Configurations in Humans
- 1 July 1994
- journal article
- clinical trial
- Published by Wiley in Journal of Cardiovascular Electrophysiology
- Vol. 5 (7) , 581-590
- https://doi.org/10.1111/j.1540-8167.1994.tb01299.x
Abstract
Biphasic Defibrillation with Nonthoracotomy Leads. Introduction: For current implantable defibrillators, the nonthoracotomy approach to implantation fails in a substantial number of patients. In a prospective randomized cross‐over study the defibrillation efficacy of a standard monophasic and a new biphasic waveform was compared for different lead configurations. Methods and Results: Intraoperatively, in 79 patients receiving nonthoracotomy defibrillation leads, the defibrillation threshold was determined in the initial lead configuration for the mono‐and biphasic waveform. In each patient, both waveforms were used alternately with declining energies (20, 15,10, 5 J) until failure of defibrillation occurred. Three different initial lead configurations were tested in different, consecutive, nonrandomized patients using a bipolar endocardial defibrillation lead alone (A; n = 36) or in combination with a subcutaneous defibrillation patch (B; n = 24) or array (C; n = 19) lead. The lowest successful defibrillation energy with the biphasic waveform was less than, equal to, or higher than with the monophasic waveform in 64%, 28%, and 8% of patients, respectively, and on average significantly lower with the biphasic waveform for all three lead configurations (A: 11.3 ± 4.4 J vs 14.5 ± 4.5.); B: 9.7 ± 4.7 J vs 15.1 ± 4.5 J; C: 7.9 ± 4.5 J vs 12.4 ± 4.9 J). Defibrillation efficacy at 20 J was significantly improved by the biphasic waveform (91% vs 76%). Conclusion: In combination with nonthoracotomy defibrillation leads, the biphasic waveform of a new implantable cardioverter defibrillator showed superior defibrillation efficacy in comparison to the standard monophasic waveform. Defibrillation thresholds were improved for lead systems with and without a subcutaneous patch or array lead.Keywords
This publication has 27 references indexed in Scilit:
- Endocardial pacing, cardioversion and defibrillation using a braided endocardial lead systemThe American Journal of Cardiology, 1993
- Comparison of biphasic and monophasic shocks for defibrillation using a nonthoracotomy systemThe American Journal of Cardiology, 1993
- Effects of chronic amiodarone therapy on defibrillation thresholdThe American Journal of Cardiology, 1992
- Efficacy and safety of monophasic and biphasic waveform shocks using a braided endocardial defibrillation lead systemAmerican Heart Journal, 1990
- A prospective randomized evaluation of biphasic versus monophasic waveform pulses on defibrillation efficacy in humansJournal of the American College of Cardiology, 1989
- Superiority of biphasic shocks in the defibrillation of dogs by epicardial patches and catheter electrodesAmerican Heart Journal, 1989
- Comparison of monophasic with single and dual capacitor biphasic waveforms for nonthoracotomy canine internal defibrillationJournal of the American College of Cardiology, 1989
- Improved internal defibrillation efficacy with a biphasic waveformAmerican Heart Journal, 1989
- Ventricular defibrillation using biphasic waveforms: The importance of phasic durationJournal of the American College of Cardiology, 1989
- Improved low energy defibrillation efficacy in man with the use of a biphasic truncated exponential waveformAmerican Heart Journal, 1989