A Limited Protocol for Assessment of Within-Batch Performance of Multichannel Analysers
- 1 January 1978
- journal article
- research article
- Published by SAGE Publications in Annals of Clinical Biochemistry: International Journal of Laboratory Medicine
- Vol. 15 (1-6) , 12-17
- https://doi.org/10.1177/000456327801500102
Abstract
An experimental protocol is described to evaluate rapidly the performance of biochemical analysers. A trial in which this protocol was applied to 15 multichannel continuous-flow analysers is described and the results are discussed. Some consideration is given to the effects of applying a commonly-used drift correction procedure to the raw data, and values of within-batch standard deviation are proposed which may be useful in assessing performance.This publication has 2 references indexed in Scilit:
- Definitions of Some Words and Terms Used in Automated AnalysisAnnals of Clinical Biochemistry: International Journal of Laboratory Medicine, 1974
- A Revised Scheme for the Evaluation of Automatic Instruments for Use in Clinical ChemistryAnnals of Clinical Biochemistry: International Journal of Laboratory Medicine, 1974