On Line Infrared Identification of Gas Chromatography Effluents
- 1 January 1975
- journal article
- research article
- Published by SAGE Publications in Applied Spectroscopy
- Vol. 29 (1) , 24-28
- https://doi.org/10.1366/000370275774455491
Abstract
On-line infrared spectra obtained with the Norcon RS-1 gas chromatographic identifier are of sufficient quality for qualitative identification of functional groups, chemical classes, or chemical entities. Identification limits for chemical entities vary between 20 and 200 µg, polar compounds having higher sensitivities. Functional group identifications can be made with correspondingly smaller samples. A 5-fold increase in sensitivity was demonstrated using signal-averaging techniques.Keywords
This publication has 6 references indexed in Scilit:
- Gas chromatograph peaks identified on-line by a new grating infrared spectrophotometerAnalytical Chemistry, 1973
- Measurements of infrared spectra of gas-liquid chromatography fractions using multiple-scan interference spectrometryAnalytical Chemistry, 1967
- A Rapid Scanning Infrared-Gas Chromatography Instrument.Analytical Chemistry, 1964
- Construction and Performance of an Infrared Chromatographic Fraction Analyzer.Analytical Chemistry, 1964
- The application of a long-path infra-red cell in gas chromatographyJournal of Chromatography A, 1963
- Organic composition analysis by combination of gas chromatography with infrared spectrophotometryAnalytica Chimica Acta, 1961