Axial Flow Ventricular Assist Device: System Performance Considerations
- 1 January 1994
- journal article
- Published by Wiley in Artificial Organs
- Vol. 18 (1) , 44-48
- https://doi.org/10.1111/j.1525-1594.1994.tb03297.x
Abstract
A cooperative effort between Baylor College of Medicine and NASA/Johnson Space Center is under way to develop an implantable left ventricular assist device for either pulmonary or systemic circulatory support for more than 3 months' duration. Using methodical evaluation and testing, an implantable axial pump has been systematically improved. These improvements include the addition of an inducer as a pumping element in front of the impeller and the construction of an efficient brushless direct current motor. To date, less than 10 W of power is required to generate 5 L/min flow against 100 mm Hg. An index of hemolysis of 0.021 g/100 L has been achieved. Two‐day in vivo feasibility studies in calves are under way to evaluate the antithrombogenic nature of the pump. Further improvements in system efficiency, hemolytic performance, and the antithrombogenic nature of the pump are expected with the use of empirical studies, computer flow modeling, and in vivo testing in calves.Keywords