An evaluation of transverse axial emission tomography of the brain in the clinical situation
- 1 August 1977
- journal article
- research article
- Published by Oxford University Press (OUP) in The British Journal of Radiology
- Vol. 50 (596) , 555-561
- https://doi.org/10.1259/0007-1285-50-596-555
Abstract
The Aberdeen section scanner provides, in addition to conventional views, a display of the distribution of radioactivity in a tomographic section of the brain. Scans of lesions confirmed pathologically or radiologically, along with normal and doubtful cases, were examined after a careful randomization procedure. During the course of the study each set of conventional views was examined twice, once without and once with a tomographic section view. In 2 of 29 cases (7%) the lesion was diagnosed only with the aid of the tomographic section scan. Improved accuracy of diagnosis with the tomographic section view was significant at the 1% level or better.This publication has 5 references indexed in Scilit: