Analysis of spherical aberration of a water immersion objective: application to specimens with refractive indices 1.33–1.40
- 1 March 2000
- journal article
- research article
- Published by Wiley in Journal of Microscopy
- Vol. 197 (3) , 274-284
- https://doi.org/10.1046/j.1365-2818.2000.00635.x
Abstract
The method of using immersion medium to correct spherical aberration for water immersion objectives when the samples are not water is investigated. Spherical aberration is measured by an interferometer converted from a confocal microscope for samples with different refractive indices. When the proper refractive index of the immersion medium and thickness of cover slip are selected, the measured spherical aberration approaches zero. A theoretical model can be used for prediction of the immersion medium to correct spherical aberration for various samples. Using the thinnest available cover slip (100 μm), the zero spherical aberration condition can be applied to samples with refractive index as high as 1.40. Confocal images in the condition of almost no spherical aberration are included to demonstrate the improvement of axial resolution due to this correction.Keywords
This publication has 7 references indexed in Scilit:
- Effects of specimen refractive index on confocal imagingJournal of Microscopy, 1997
- Confocal optical microscopyReports on Progress in Physics, 1996
- Determination of the refractive index of highly scattering human tissue by optical coherence tomographyOptics Letters, 1995
- In Vivo Confocal Scanning Laser Microscopy of Human Skin: Melanin Provides Strong ContrastJournal of Investigative Dermatology, 1995
- Objective Lenses for Confocal MicroscopyPublished by Springer Nature ,1995
- Aberrations in confocal fluorescence microscopy induced by mismatches in refractive indexJournal of Microscopy, 1993
- Axial imaging through an aberrating layer of water in confocal microscopyOptics Communications, 1992