MEASUREMENT OF CARBON DIOXIDE OUTPUT, ALVEOLAR CARBON DIOXIDE CONCENTRATION AND ALVEOLAR VENTILATION IN THE NEONATE
Open Access
- 1 May 1977
- journal article
- research article
- Published by Elsevier in British Journal of Anaesthesia
- Vol. 49 (5) , 453-456
- https://doi.org/10.1093/bja/49.5.453
Abstract
A system is described for measuring carbon dioxide output and alveolar carbon dioxide concentration in the neonate and for deriving the volume of alveolar ventilation. Air is drawn through a soft rubber face mask applied loosely to the face so that all expired gas is entrained. A fraction of the flow is sampled by an infra-red analyser, the signal from which is integrated to provide a measure of carbon dioxide output. Alveolar carbon dioxide concentration is measured by continuous sampling of expired gas through a nasal catheter. The response time of the catheter-analyser system is 140 ms. In 20 healthy 1-day-old neonates the mean carbon dioxide excretion was 5.86 ml/min.kg-1(STPD) and the mean alveolar Pco2 was 4.85 kPa. From these measurements mean alveolar ventilation 147ml/min.kg-1 (BTPS) was calculated. These results are similar to those obtained by other methods.This publication has 0 references indexed in Scilit: