Calibration of a practical haemoglobinometer
- 1 December 1987
- journal article
- Published by Wiley in Clinical and Laboratory Haematology
- Vol. 9 (4) , 387-393
- https://doi.org/10.1111/j.1365-2257.1987.tb00576.x
Abstract
A practical instrument for routine haemoglobinometry (HemoCue) was compared with the reference haemiglobincyanide (methaemoglobincyanide (HiCN] method. The precision of the routine method was adequate, but a systematic error of about -3.5% was found. As this resulted from errors in the procedure for coupling the haemoglobinometer read-out to the reference HiCN method, we drew up an improved calibration procedure for practical haemoglobinometers to ensure comparability of results with the reference method. After recalibration according to these rules, the HemoCue system proved to be accurate to within +/- 1.5%.Keywords
This publication has 4 references indexed in Scilit:
- Spectrophotometry of Hemoglobin and Hemoglobin DerivativesPublished by Elsevier ,1983
- International Committee for Standardization in Haematology. Recommendations for reference method for haemoglobinometry in human blood (ICSH standard EP 6/2: 1977) and specifications for international haemiglobincyanide reference preparation (ICSH standard EP 6/3: 1977).Journal of Clinical Pathology, 1978
- Standardization of hemoglobinometry II. The hemiglobincyanide methodClinica Chimica Acta; International Journal of Clinical Chemistry, 1961