Beam characteristics of the Therapax DXT300 orthovoltage therapy unit
- 1 September 1994
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 39 (9) , 1377-1392
- https://doi.org/10.1088/0031-9155/39/9/006
Abstract
The operating performance and beam characteristics of a new orthovoltage unit, the Therapax DXT300, have been evaluated. Percentage depth-dose and backscatter tables are presented for several applicator sizes, at 30 cm and 50 cm focal skin distances (FSDS) and for multiple X-ray beam qualities with the tube operating between 100 and 300 kVp accelerating potential. The unit has been found to provide beam characteristics similar to those reported for other orthovoltage therapy machines. The linearity and short- and longterm stability/reproducibility of the unit's internal dosimetry system have also been studied, and results indicate a very stable beam output of better than 1% standard deviation. The data presented in this work should provide the basis for comparison with other units and act as a reference for clinics commissioning the Therapax DXT300 in the future.Keywords
This publication has 8 references indexed in Scilit:
- Differences in reported backscatter factors for low-energy X-rays: a literature studyPhysics in Medicine & Biology, 1993
- Dependence of the photon backscatter factor for water on irradiation field size and source-to-phantom distances between 1.5 and 10 cmPhysics in Medicine & Biology, 1993
- Dosimetry of a kilovoltage radiotherapy x-ray machineMedical Dosimetry, 1993
- Report on the IPSM working party on low- and medium-energy x-ray dosimetryPhysics in Medicine & Biology, 1991
- Dependence of the photon backscatter factor for water on source-to-phantom distance and irradiation field sizePhysics in Medicine & Biology, 1990
- Experimentally determined backscatter factors for X-rays generated at voltages between 16 and 140 kVPhysics in Medicine & Biology, 1989
- Performance characteristics of an orthovoltage x‐ray therapy machineMedical Physics, 1987
- Performance characteristics of a widely used orthovoltage x‐ray therapy machineMedical Physics, 1986