In the first part of the review the various mechanisms that could be the cause of interference in bioelectric recordings are considered. It is demonstrated that the performance of a good amplifier can be seriously degraded in its functioning if the whole measurement situation is not taken into account. Several techniques used to reduce interference, of which guarding and driven right leg circuits are the most important, are analysed. In the second part of the review some examples of the application of the theory in practical situations are presented. The instrumentation amplifier circuit normally used in bioelectric recordings is improved for measurements under difficult circumstances. Another application is a low-cost 64-channel amplifier for multichannel ECG recordings. The third application is a device that can be added to bioelectric measurement systems and will provide a major reduction in interference.Keywords
This publication has 17 references indexed in Scilit:
- Radiotransparent carbon electrode for ECG recordings in the catheterization laboratoryIEEE Transactions on Biomedical Engineering, 1988
- Interference-rejection characteristics of biopotential amplifiers: a comparative analysisIEEE Transactions on Biomedical Engineering, 1988
- Recording of surface His-Purkinje potentialsMedical & Biological Engineering & Computing, 1985
- Low-power, low-noise instrumentation amplifier for physiological signalsMedical & Biological Engineering & Computing, 1984
- Impedance measurement of individual skin surface electrodesMedical & Biological Engineering & Computing, 1983
- Driven-right-leg circuit designIEEE Transactions on Biomedical Engineering, 1983
- Measurement of patient body capacitance and a method of patient isolation in mains environmentsMedical & Biological Engineering & Computing, 1974
- 60-Hz Interference in ElectrocardiographyIEEE Transactions on Biomedical Engineering, 1973
- SYSTEMIC AND RANDOM VARIATIONS OF ECG ELECTRODE SYSTEM IMPEDANCE*Annals of the New York Academy of Sciences, 1970