Progress in multipass tandem Fabry–Perot interferometry: I. A fully automated, easy to use, self-aligning spectrometer with increased stability and flexibility
- 1 March 1999
- journal article
- conference paper
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 70 (3) , 1589-1598
- https://doi.org/10.1063/1.1149637
Abstract
A computer control for a Sandercock-type multipath tandem Fabry–Perot interferometer is described, which offers many advantages over conventionally used analog control: The range of stability is increased due to active control of the laser light intensity and the mirror dither amplitude. The alignment is fully automated enabling start of a measurement within a minute after start of align, including optionally finding the optimum focus on the sample. The software control enables a programmable series of measurements with control of, e.g., the position and rotation of the sample, the angle of light incidence, the sample temperature, or the strength and direction of an applied magnetic field. Built-in fitting routines allow for a precise determination of frequency positions of excitation peaks combined with increased frequency accuracy due to a correction of a residual nonlinearity of the mirror stage drive.Keywords
This publication has 21 references indexed in Scilit:
- A new design for a versatile Fabry-Pérot interferometer for Brillouin spectroscopyReview of Scientific Instruments, 1997
- Construction of a Vernier Tandem (3+3)-Pass Brillouin Spectrometer and Its Application to Surface Acoustic Wave Study in Metallic Multilayer FilmsJapanese Journal of Applied Physics, 1994
- Magnetic anisotropies of ultrathin Co(001) films on Cu(001)Physical Review Letters, 1992
- A Dynamic Antivibration SystemPublished by SPIE-Intl Soc Optical Eng ,1987
- Construction and performance of a Brillouin scattering set-up using a triple-pass tandem Fabry-Perot interferometerJournal of Physics E: Scientific Instruments, 1987
- Construction and alignment of a high performance multipass vernier tandem Fabry–Perot interferometerReview of Scientific Instruments, 1981
- Tandem multipass Fabry-Perot interferometer for Brillouin scatteringApplied Optics, 1981
- Light scattering from surface acoustic phonons in metals and semiconductorsSolid State Communications, 1978
- Linear scanning circuit for a piezoelectrically controlled Fabry–Perot etalonReview of Scientific Instruments, 1977
- Digital stabilizer for scanning Fabry–PerotReview of Scientific Instruments, 1976