Pre-metrological (Pre-analytical) Variation of Some Biochemical Quantities
- 15 October 1999
- journal article
- Published by Walter de Gruyter GmbH in cclm
- Vol. 37 (10) , 987-989
- https://doi.org/10.1515/cclm.1999.146
Abstract
It is customary to assume that pre-metrological variation is negligible if all pre-metrological sources of variation have been standardised. The aim of this study was to quantify the pre-metrological variation of some biochemical quantities frequently measured in clinical laboratories. The experimental design considers the different sources of pre-metrological variation that usually act day-to-day on samples from patients. The results demonstrate that, for many quantities, the pre-metrological variance is not negligible; consequently, it is suggested that these variances should be taken into account when estimating the uncertainty of measurement of patients' results.Keywords
This publication has 1 reference indexed in Scilit:
- Generation and Application of Data on Biological Variation in Clinical ChemistryCritical Reviews in Clinical Laboratory Sciences, 1989