Radiation dose reduction in paediatric fluoroscopy using added filtration
- 1 March 1995
- journal article
- Published by Oxford University Press (OUP) in The British Journal of Radiology
- Vol. 68 (807) , 296-300
- https://doi.org/10.1259/0007-1285-68-807-296
Abstract
Studies on a water phantom indicate that the insertion of extra tube filtration combined with the removal of the antiscatter grid can reduce the total energy imparted from paediatric fluoroscopy by a factor of more than four when automatic exposure control is used. The omission of the grid is responsible for up to 40% of the reduction while the addition of 0.7 mm steel filtration is found to account for a further reduction in energy imparted of between 57% and 70%, for depths of water simulating small patients of 6-16 cm thickness. For all contrast paediatric examinations the reduction in image quality was considered to be small and the diagnostic information unimpaired with the filter in place.Keywords
This publication has 6 references indexed in Scilit:
- Radiation doses to paediatric patients undergoing less common radiological procedures involving fluoroscopyThe British Journal of Radiology, 1993
- Results of a survey of doses to paediatric patients undergoing common radiological examinationsThe British Journal of Radiology, 1992
- Dose reduction by the use of erbium filtration in a general radiographic roomThe British Journal of Radiology, 1992
- The reduction in radiation dose for intra-oral radiographs by the use of thin K-edge filtersThe British Journal of Radiology, 1991
- Application of added beam filtration to intra-oral radiography.Dento maxillo facial radiology, 1990
- Fluoroscopy without the grid: a method of reducing the radiation doseThe British Journal of Radiology, 1980