Newer Portable Glucose Meters—Analytical Improvement Compared with Previous Generation Devices?
- 1 October 1999
- journal article
- research article
- Published by Oxford University Press (OUP) in Clinical Chemistry
- Vol. 45 (10) , 1821-1825
- https://doi.org/10.1093/clinchem/45.10.1821
Abstract
Background: Newer glucose meters are easier to use, but direct comparisons with older instruments are lacking. We wished to compare analytical performances of four new and four previous generation meters. Methods: On average, 248 glucose measurements were performed with two of each brand of meter on capillary blood samples from diabetic patients attending our outpatient clinic. Two to three different lots of strips were used. All measurements were performed by one experienced technician, using blood from the same sample for the meters and the comparison method (Beckman Analyzer 2). Results were evaluated by analysis of clinical relevance using the percentage of values within a maximum deviation of 5% from the reference value, by the method of residuals, by error grid analysis, and by the CVs for measurements in series. Results: Altogether, 1987 blood glucose values were obtained with meters compared with the reference values. By error grid analysis, the newer devices gave more accurate results without significant differences within the group (zone A, 98–98.5%). Except for the One Touch II (zone A, 98.5%), the other older devices were less exact (zone A, 87–92.5%), which was also true for all other evaluation procedures. Conclusions: New generation blood glucose meters are not only smaller and more aesthetically appealing but are more accurate compared with previous generation devices except the One Touch II. The performance of the newer meters improved but did not meet the goals of the latest American Diabetes Association recommendations in the hands of an experienced operator.This publication has 11 references indexed in Scilit:
- Spectroscopic and Clinical Aspects of Noninvasive Glucose MeasurementsClinical Chemistry, 1999
- Clinical and statistical evaluation of self-monitoring blood glucose meters.Diabetes Care, 1998
- Intensive blood-glucose control with sulphonylureas or insulin compared with conventional treatment and risk of complications in patients with type 2 diabetes (UKPDS 33)The Lancet, 1998
- Validation of Home Blood Glucose Meters With Respect to Clinical and Analytical ApproachesDiabetes Care, 1998
- Understanding Error Grid AnalysisDiabetes Care, 1997
- Accuracy, precision and user-acceptability of self blood glucose monitoring machinesDiabetes Research and Clinical Practice, 1997
- Noninvasive Blood Glucose MonitoringDiabetes Care, 1997
- What Do Statistics Really Tell Us About the Quality of the Data from Self-Monitoring of Blood Glucose?Diabetic Medicine, 1989
- Evaluating Clinical Accuracy of Systems for Self-Monitoring of Blood GlucoseDiabetes Care, 1987
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986