Performance characteristics of the 3-D OSEM algorithm in the reconstruction of small animal PET images
- 1 January 2000
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Medical Imaging
- Vol. 19 (8) , 798-804
- https://doi.org/10.1109/42.876305
Abstract
Rat brain images acquired with a small animal positron emission tomography (PET) camera and reconstructed with the three-dimensional (3-D) ordered-subsets expectation-maximization (OSEM) algorithm with resolution recovery have better quality when the brain is imaged by itself than when inside the head with surrounding background activity. The purpose of this study was to characterize the dependence of this effect on the level of background activity, attenuation, and scatter. Monte Carlo simulations of the imaging system were performed. The coefficient of variation from replicate images, full-width at half-maximum (FWHM) from point sources and image profile fitting, and image contrast and uniformity were used to evaluate algorithm performance. A rat head with the typical levels of five and ten times the brain activity in the surrounding background requires additional iterations to achieve the same resolution as the brain-only case at a cost of 24% and 64% additional noise, respectively. For the same phantoms, object scatter reduced contrast by 3%-5%. However, attenuation degraded resolution by 0.2 mm and was responsible for up to 12% nonuniformity in the brain images suggesting that attenuation correction is useful. Given the effects of emission and attenuation distribution on both resolution and noise, simulations or phantom studies should be used for each imaging situation to select the appropriate number of OSEM iterations to achieve the desired resolution-noise levels.Keywords
This publication has 20 references indexed in Scilit:
- Development of Fluorine-18-Labeled 5-HT1AAntagonistsJournal of Medicinal Chemistry, 1999
- Positron emission tomography (pet) methodology for small animals and its application in radiopharmaceutical preclinical investigationNuclear Medicine and Biology, 1998
- An evaluation of exact and approximate 3-D reconstruction algorithms for a high-resolution, small-animal PET scannerIEEE Transactions on Medical Imaging, 1998
- Evaluation of 3D reconstruction algorithms for a small animal PET cameraIEEE Transactions on Nuclear Science, 1997
- The design and physical characteristics of a small animal positron emission tomographPhysics in Medicine & Biology, 1995
- Developments in tomographic methods for biological trace element researchBiological Trace Element Research, 1994
- Accelerated image reconstruction using ordered subsets of projection dataIEEE Transactions on Medical Imaging, 1994
- The convergence of object dependent resolution in maximum likelihood based tomographic image reconstructionPhysics in Medicine & Biology, 1993
- Analytic 3D image reconstruction using all detected eventsIEEE Transactions on Nuclear Science, 1989
- Object-dependent performance comparison of two iterative reconstruction algorithmsIEEE Transactions on Nuclear Science, 1988