Factors affecting image quality and diagnostic efficacy in abdominal sonography: A prospective study of 140 patients
- 1 November 1993
- journal article
- research article
- Published by Wiley in Journal of Clinical Ultrasound
- Vol. 21 (9) , 623-630
- https://doi.org/10.1002/jcu.1870210909
Abstract
A study was undertaken to evaluate the limitations of abdominal sonography in a group of predominantly elderly patients. In the majority of patients (98%) sonography, using 3.5 MHz and 5 MHz array transducers permitted visualization of the major abdominal organs with image quality sufficient for definitive diagnosis, but improved image quality would be desirable in 78% of this sample. Obesity and immobilization were associated with poor sonographic image quality. In 31 of the 45 (69%) poor image quality studies, no near‐field artifacts (less than 3 cm depth) were observed. Obstacles in the acoustic path were noted in 77% of the poor image quality studies. © 1993 John Wiley & Sons, Inc.Keywords
This publication has 10 references indexed in Scilit:
- Direct estimation of aberrating delays in pulse-echo imaging systemsThe Journal of the Acoustical Society of America, 1990
- A sonographic technique to reduce beam distortion by curved interfacesUltrasound in Medicine & Biology, 1989
- Phase aberration measurements in medical ultrasound: Human studiesUltrasonic Imaging, 1988
- Phase Aberration Measurements in Medical Ultrasound: Human StudiesUltrasonic Imaging, 1988
- An iterative phase correction scheme based upon speckle brightnessUltrasonic Imaging, 1988
- Considerations in the use of echocardiography in epidemiology. The Framingham Study.Hypertension, 1987
- The ghost artifact.Journal of Ultrasound in Medicine, 1984
- The importance of ultrasonic side-lobe artifacts.Radiology, 1982
- Slice-thickness artifacts in gray-scale ultrasoundJournal of Clinical Ultrasound, 1981
- Range ambiguities in real‐time ultrasoundJournal of Clinical Ultrasound, 1981