Comparison of three keratometry instruments
- 1 July 1995
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of Cataract & Refractive Surgery
- Vol. 21 (4) , 407-408
- https://doi.org/10.1016/s0886-3350(13)80530-5
Abstract
We compared the reliability of measurements of three keratometers and keratometry readings from a corneal topography system to determine if they were as accurate as the "gold standard" Javal-Schiotz keratometer. Same-day measurement of the steepest and flattest powers of the central 3 mm of the corneas of 200 eyes (100 patients) revealed no statistically significant difference in reliability between the newer keratometers and the Javal-Schiotz. In addition, the newer machines are more convenient and efficient in some clinical settings.Keywords
This publication has 2 references indexed in Scilit:
- Comparison of the accuracy and reproducibility of the keratometer and the EyeSys Corneal Analysis System Model IJournal of Cataract & Refractive Surgery, 1992
- Clinical comparison of automated and manual keratometry in pre-operative ocular biometryEye, 1992