Temperature-Insensitive Near-Infrared Spectroscopic Measurement of Glucose in Aqueous Solutions
- 1 April 1994
- journal article
- Published by SAGE Publications in Applied Spectroscopy
- Vol. 48 (4) , 477-483
- https://doi.org/10.1366/000370294775268910
Abstract
A digital Fourier filter is combined with partial least-squares (PLS) regression to generate a calibration model for glucose that is insensitive to sample temperature. This model is initially created by using spectra collected over the 5000 to 4000 cm−1 spectral range with samples maintained at 37°C. The analytical utility of the model is evaluated by judging the ability to determine glucose concentrations from a set of prediction spectra. Absorption spectra in this prediction set are obtained by ratioing single-beam spectra collected from solutions at temperatures ranging from 32 to 41°C to reference spectra collected at 37°C. The temperature sensitivity of the underlying water absorption bands creates large baseline variations in prediction spectra that are effectively eliminated by the Fourier filtering step. The best model provides a mean standard error of prediction across temperatures of 0.14 mM (2.52 mg/dL). The benefits of the Fourier filtering step are established, and critical experimental parameters, such as number of PLS factors, mean and standard deviation for the Gaussian shaped Fourier filter, and spectral range, are considered.Keywords
This publication has 8 references indexed in Scilit:
- Near-infrared spectroscopic measurement of glucose in a protein matrixAnalytical Chemistry, 1993
- Strategies for coupling digital filtering with partial least-squares regression: Application to the determination of glucose in plasma by Fourier-transform near-infrared spectroscopyAnalytical Chemistry, 1993
- Near-Infrared Spectrophotometry: A New Dimension in Clinical ChemistryClinical Chemistry, 1992
- Understanding spectra of highly excited vibrational statesPublished by SPIE-Intl Soc Optical Eng ,1992
- Near-infrared spectroscopy: the new FT frontier--spectroscopic and chemometric considerationsPublished by SPIE-Intl Soc Optical Eng ,1992
- Determination of physiological levels of glucose in an aqueous matrix with digitally filtered Fourier transform near-infrared spectraAnalytical Chemistry, 1990
- Monitoring of glucose in biological fluids by fourier-transform infrared spectrometry with a cylindrical internal reflectance cellAnalytica Chimica Acta, 1987