Optimum adjustment and correction of the Wien filter
- 1 November 1967
- journal article
- Published by IOP Publishing in British Journal of Applied Physics
- Vol. 18 (11) , 1573-1579
- https://doi.org/10.1088/0508-3443/18/11/309
Abstract
The crossed homogeneous electrostatic and magnetic deflection fields of an analyser (Wien filter) have been formed by two condenser plates and two pairs of coils. The optimum operating conditions are shown to require unit magnification, contrary to the view of other authors. By varying the current ratio in the coils, a desired inhomogeneity can be introduced in the magnetic field and with this the second-order angular aberration can be corrected. The possibility of a double-focusing Wien filter is discussed.Keywords
This publication has 7 references indexed in Scilit:
- Das Auflösungsvermögen des elektrostatisch-magnetischen Energieanalysators für schnelle ElektronenThe European Physical Journal A, 1964
- Ein modifiziertes Wiensches Filter als ElektronenmonochromatorThe European Physical Journal A, 1963
- Bestätigung der anomalen Energiebreiten in Elektronenstrahlen mit dem elektrostatischen AnalysatorThe European Physical Journal A, 1958
- Experimentelle Bestimmung der Energieverteilung in thermisch ausgelösten ElektronenstrahlenThe European Physical Journal A, 1954
- Angular Aberrations in Sector Shaped Electromagnetic Lenses for Focusing Beams of Charged ParticlesPhysical Review B, 1953
- Design Curves for 180° Magnetic SpectrometersReview of Scientific Instruments, 1949
- Feldkombinationen zur Geschwindigkeits‐ und Massenspektrographie. IAnnalen der Physik, 1934