Determinations of N gas and P repl factors from commercially available parallel-plate chambers: AAPM Task Group 21 protocol
- 1 November 1986
- journal article
- Published by Wiley in Medical Physics
- Vol. 13 (6) , 908-912
- https://doi.org/10.1118/1.595968
Abstract
Using a Farmer chamber as a reference dosimeter, we have measured the Ngas (cavity-gas calibration factor) and Prepl (replacement correction factor) values for four parallel-plate chambers: a Holt chamber, a Capintec chamber, a Markus chamber, and an SHM chamber. Ngas determinations were done primarily using 60Co photons and nominal 18-MeV electrons for Ngas determinations. For Prepl determinations, we used nominal 18-, 15-, 12-, 9-, 6-, and 4-MeV electrons.Keywords
This publication has 0 references indexed in Scilit: