Thermal dye double indicator dilution measurement of lung water in man: comparison with gravimetric measurements.
- 1 January 1987
- Vol. 42 (1) , 72-76
- https://doi.org/10.1136/thx.42.1.72
Abstract
The thermal dye double indicator dilution technique for estimating lung water was compared with gravimetric analyses in nine human subjects who were organ donors. As observed in animal studies, the thermal dye measurement of extravascular thermal volume (EVTV) consistently overestimated gravimetric extravascular lung water (EVLW), the mean (SEM) difference being 3.43 (0.59) ml/kg. In eight of the nine subjects the EVTV - 3.43 ml/kg would yield an estimate of EVLW that would be from 3.23 ml/kg under to 3.37 ml/kg over the actual value EVLW at the 95% confidence limits. Reproducibility, assessed with the standard error of the mean percentage, suggested that a 15% change in EVTV can be reliably detected with repeated measurements. One subject was excluded from analysis because the EVTV measurement grossly underestimated its actual EVLW. This error was associated with regional injury observed on gross examination of the lung. Experimental and clinical evidence suggest that the thermal dye measurement provides a reliable estimate of lung water in diffuse pulmonary oedema states.This publication has 17 references indexed in Scilit:
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- Evaluation of the Portable Chest Roentgenogram for Quantitating Extravascular Lung Water in Critically Ill AdultsChest, 1985
- Effect of graded administration of PEEP on lung water in noncardiogenic pulmonary edemaCritical Care Medicine, 1982
- Measurement of Extravascular Lung Water in Dogs Using the Thermal-Green Dye Indicator Dilution MethodAnesthesiology, 1982
- THE MEASUREMENT OF EXTRAVASCULAR LUNG WATER BY THERMAL‐GREEN DYE INDICATOR DILUTION*Annals of the New York Academy of Sciences, 1982
- RELIABILITY OF THE THERMODILUTION METHOD IN THE DETERMINATION OF CARDIAC-OUTPUT IN CLINICAL-PRACTICEPublished by Elsevier ,1982
- Evaluation of transthoracic electrical impedance in the diagnosis of pulmonary edema.Circulation, 1979
- Bedside measurement of lung waterJournal of Surgical Research, 1979
- Thermal-dye lung water measurements: Effects of edema and embolizationJournal of Surgical Research, 1979
- Excessive fluid administration in resuscitating baboons from hemorrhagic shock, and an assessment of the thermodye technic for measuring extravascular lung waterThe American Journal of Surgery, 1978