MISR radiometric uncertainty analyses and their utilization within geophysical retrievals
- 1 August 1998
- journal article
- Published by IOP Publishing in Metrologia
- Vol. 35 (4) , 571-579
- https://doi.org/10.1088/0026-1394/35/4/61
Abstract
The Multi-angle Imaging SpectroRadiometer (MISR) is to be launched with the Earth Observing System EOS-AM1 spacecraft in 1999. Demanding specifications include a requirement that the instrument be calibrated, and placed on an accurate radiometric scale, to within 3 % uncertainty, for incident radiances near the upper end of a camera's dynamic range. Contributing factors to the uncertainty in radiance measurements include signal-to-noise ratio, goodness of fit of the calibration data to a quadratic equation, and quality of the experimental conditions (i.e. range and number of radiometric levels used to provide the calibration). These sources, as they are identified, are flagged as contributing to the absolute, relative band-to-band, relative camera-to-camera, or relative pixel-to-pixel radiance uncertainties. This paper summarizes the products and results from the uncertainty analysis of radiometric calibrations for the MISR. It also provides an example of a geophysical product retrieval that makes use of these data.Keywords
This publication has 9 references indexed in Scilit:
- Techniques for the retrieval of aerosol properties over land and ocean using multiangle imagingIEEE Transactions on Geoscience and Remote Sensing, 1998
- Determination of land and ocean reflective, radiative, and biophysical properties using multiangle imagingIEEE Transactions on Geoscience and Remote Sensing, 1998
- Multi-angle Imaging SpectroRadiometer (MISR) instrument description and experiment overviewIEEE Transactions on Geoscience and Remote Sensing, 1998
- MISR prelaunch instrument calibration and characterization resultsIEEE Transactions on Geoscience and Remote Sensing, 1998
- The Airborne Multi-angle Imaging SpectroRadiometer (AirMISR): instrument description and first resultsIEEE Transactions on Geoscience and Remote Sensing, 1998
- Estimation of calibration uncertainties for the Multiangle Imaging SpectroRadiometer (MISR) via fidelity intervalsPublished by SPIE-Intl Soc Optical Eng ,1993
- Calibration Plans for the Multi-angle Imaging SpectroRadiometer (MISR)Metrologia, 1993
- Comparison of two cryogenic radiometers by determining the absolute spectral responsivity of silicon photodiodes with an uncertainty of 002%Applied Optics, 1990
- Spectral response self-calibration and interpolation of silicon photodiodesApplied Optics, 1980