Interlaboratory validation of a CD71-based flow cytometric method (Microflow®) for the scoring of micronucleated reticulocytes in mouse peripheral blood
- 16 December 2004
- journal article
- research article
- Published by Wiley in Environmental and Molecular Mutagenesis
- Vol. 45 (1) , 44-55
- https://doi.org/10.1002/em.20081
Abstract
An interlaboratory study was performed to validate an anti‐CD71/flow cytometry‐based technique for enumerating micronucleated reticulocytes (MN‐RETs) in mouse peripheral blood. These experiments were designed to address International Workshop on Genotoxicity Test Procedures validation criteria by evaluating the degree of correspondence between MN‐RET measurements generated by flow cytometry (FCM) with those obtained using traditional microscopy‐based methods. In addition to these cross‐methods data, flow cytometric MN‐RET measurements for each blood sample were performed at two separate sites in order to evaluate the reproducibility of data between laboratories. In these studies, groups of male CD‐1 mice were treated with vehicle (saline or vegetable oil), a negative control (saline or vegetable oil), or four dose levels of five known genotoxicants (clastogens: cyclophosphamide, benzo[a]pyrene, 5‐fluorouracil, methotrexate; aneugen: vincristine sulfate). Exposure occurred on 3 consecutive days via intraperitoneal injection, and blood samples were obtained approximately 24 hr after the final treatment. MN‐RET frequencies were determined for each sample based on the analysis of 2,000 (microscopy) and 20,000 (FCM) reticulocytes. Regardless of the method utilized, each genotoxic agent was observed to cause statistically significant increases in the frequency of MN‐RETs, and each response occurred in a dose‐dependent manner. Spearman's correlation coefficient (rs) for FCM versus microscopy‐based MN‐RET measurements (nine experiments, 252 paired measurements) was 0.740, indicating a high degree of correspondence between methods. The rs value for all flow cytometric MN‐RET measurements performed at the two independent sites was 0.857 (n = 248), suggesting that the automated method is highly transferable between laboratories. Additionally, the flow cytometric system offered advantages relative to microscopy‐based scoring, including a greater number of cells analyzed, much faster analysis times, and a greater degree of objectivity. Collectively, data presented in this report suggest that the overall performance of mouse peripheral blood micronucleus tests is enhanced by the use of the flow cytometric scoring procedure. Environ. Mol. Mutagen., 2005.Keywords
Funding Information
- the National Institute of Environmental Health Sciences (5R44ES11244-03)
This publication has 22 references indexed in Scilit:
- Comparative Scoring of Micronucleated Reticulocytes in Rat Peripheral Blood by Flow Cytometry and MicroscopyToxicological Sciences, 2003
- In vivo rodent erythrocyte micronucleus assay. II. Some aspects of protocol design including repeated treatments, integration with toxicity testing, and automated scoringEnvironmental and Molecular Mutagenesis, 2000
- Simple and reliable enumeration of micronucleated reticulocytes with a single-laser flow cytometerMutation Research/Genetic Toxicology, 1996
- Mouse bone marrow micronucleus test using flow cytometryMutagenesis, 1992
- Flow‐cytometric enumeration of micronucleated polychromatic erythrocytes in mouse peripheral bloodCytometry, 1992
- Flow cytometric micronucleus test with mouse peripheral erythrocytesMutagenesis, 1992
- The micronucleus assay with mouse peripheral blood reticulocytes using acridine orange-coated slidesMutation Research Letters, 1990
- An application of Acridine Orange fluorescent staining to the micronucleus testMutation Research Letters, 1983
- Rapid detection of mutagen induced micronucleated erythrocytes by flow cytometryHistochemistry and Cell Biology, 1982
- A rapid in vivo test for chromosomal damage☆Mutation Research - Fundamental and Molecular Mechanisms of Mutagenesis, 1973