Calorimetric Determination of the Energy Flux of 22.5-Mev X-Rays
- 1 August 1951
- journal article
- research article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 22 (8) , 572-574
- https://doi.org/10.1063/1.1746007
Abstract
A sensitive method of calorimetrically measuring the energy flux of an x‐ray beam is described. Thermistors are employed as the temperature‐sensitive elements and are embedded in the irradiated material in a calorimeter. Details and results of the application of this technique to a betatron x‐ray beam are described. This method can be applied to x‐ray beams of other energies and does not require high intensities.Keywords
This publication has 3 references indexed in Scilit:
- Gamma-Neutron Cross SectionsPhysical Review B, 1950
- Photo-Disintegration of Silver and AluminumPhysical Review B, 1950
- Procedure in Dose Distribution Measurement of 25-Mev X-RaysScience, 1950