Pulse oximetry versus measured arterial oxygen saturation: A comparison of the Nellcor N100 and the Biox III
- 1 January 1991
- journal article
- research article
- Published by Wiley in Pediatric Pulmonology
- Vol. 10 (2) , 132-135
- https://doi.org/10.1002/ppul.1950100216
Abstract
Pulse oximetry is noninvasive, fast, and simple, making it a very popular way of assessing oxygenation in pediatric patients. However, there are few studies that establish the accuracy of this technology over a wide range of oxygen saturations in children. This study, done in 47 children aged from 1 day to 16 years with congenital heart disease and undergoing cardiac catheterization, compared the direct measurement of arterial oxygen saturation to values from pulse oximetry. Oxygen saturation was measured by an IL‐282 Co‐oximeter, which also measured carboxyhemoglobin and methemoglobin, and was compared to values obtained from both a Biox III and Nellcor N100. Both pulse oximeters gave values that closely correlated with the actual saturation (r = 0.91 and 0.93, respectively) with standard errors of the estimate of 4.1 and 3.2%, respectively. For both devices, the error increased with decreasing saturations, being progressively larger below a saturation of 80%. The difference between the actual saturation and that measured by pulse oximetry bore no relationship to the presence of carboxyhemoglobin, methemoglobin, fetal hemoglobin, bilirubin, cardiac index, or age of the patient. In conclusion, pulse oximetry, while a very useful technology in pediatrics, must be interpreted with some caution in children with severe cyanosis. Pediatr Pulmonol 1991; 10:132–135.Keywords
This publication has 13 references indexed in Scilit:
- Accuracy of two wavelength pulse oximetry in neonates and infantsPediatric Pulmonology, 1989
- Reliability of pulse oximetry in hypoxic infantsThe Journal of Pediatrics, 1988
- NONINVASIVE PULSE OXIMETRY IN CHILDREN WITH CYANOTIC CONGENITAL HEART DISEASECritical Care Medicine, 1987
- Accuracy of Response of Six Pulse Oximeters to Profound HypoxiaAnesthesiology, 1987
- Pulse oximetry for continuous oxygen monitoring in sick newborn infantsThe Journal of Pediatrics, 1986
- Evaluation of noninvasive measurements of oxygenation in stable infantsCritical Care Medicine, 1986
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- Pulse oximetry in pediatric intensive care: Comparison with measured saturations and transcutaneous oxygen tensionThe Journal of Pediatrics, 1985
- Noninvasive arterial hemoglobin oxygen saturation versus transcutaneous oxygen tension monitoring in the preterm infantCritical Care Medicine, 1984
- Separation of haemoglobins using a monodisperse cation exchangerJournal of Chromatography A, 1984