Blood Velocity Measurements in Human Retinal Vessels
- 29 November 1974
- journal article
- other
- Published by American Association for the Advancement of Science (AAAS) in Science
- Vol. 186 (4166) , 830-831
- https://doi.org/10.1126/science.186.4166.830
Abstract
Laser Doppler velocimetry was used to measure the velocity of blood in human retinal vessels. The mean flow velocities obtained were 1.9 centimeters per second in a retinal vein and 2.2 centimeters per second in a retinal artery. Scattered light from a weak helium-neon laser beam focused on the vessel was detected by a photomultiplier, and the temporal correlation of the intensity fluctuations was measured with a photon counting autocorrelator. Autocorrelation functions for blood flowing through glass capillaries were used for calibration.Keywords