A Comparative Study of Activity and Dual Sensor: Activity and Minute Ventilation Pacing Responses to Ascending and Descending Stairs
- 1 October 1998
- journal article
- research article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 21 (10) , 1862-1868
- https://doi.org/10.1111/j.1540-8159.1998.tb00003.x
Abstract
Previous studies with activity‐based rate; adaptive pacemakers have shown a somewhat paradoxical response when comparing ascending stairs to descending stairs. The objective of this investigation was to measure dual‐sensor rate response provided by activity and minute ventilation (MV) compared with activity alone, and with a control group, during ascending and descending stairs. For dual sensor mode, measured mean peak pacing rate with 72 (92) steps per minute was 111 ± 13 beats/min (124 ± 14 beats/min) ascending stairs and 81 ± 7 beats/mm (97 ± 13 beats/min) for descending. For activity mode alone, mean peak pacing rate was 90 ± 12 beats/min (108 ± 19 beats/min) ascending stairs and 97 ± 12 beats/min (123 ±17 beats/min) descending. The mean peak control group heart rate ascending stairs for a step rate of 72 (92J steps/min were 116 ± 11 beats/min (127 ±14 beats/min) ascending stairs and for descending 89 ± 12 beats/min (95 ± 11 beats/min). While for dual sensor controlled pacing there was a significant difference for ascending and descending stairs at both step rates, there was no difference between going upstairs and downstairs for activity mode alone. Rates with dual sensor did not significantly differ from respective rates of the control group. The mean correlation coefficient between MV and paced rate was 0.85. Pacing heart rates delivered by the dual sensor mode were appropriate for ascending and descending stairs. In contrast to activity mode alone, the peak heart rates for dual sensor mode are higher during ascending than during descending stairs.Keywords
This publication has 19 references indexed in Scilit:
- Heart rate to work rate relation throughout peak exercise in normal subjects as a guideline for rate-adaptive pacemaker programmingThe American Journal of Cardiology, 1995
- Behavior of Different Activity-Based Pacemakers During Treadmill Exercise Testing with Variable Slopes: A Comparison of Three Activity-Based Pacing SystemsPacing and Clinical Electrophysiology, 1994
- ABSTRACT SESSION 9: EVALUATION OF HATE ADAPTIVE PACING SYSTEMSPacing and Clinical Electrophysiology, 1994
- Ventilation and Heart Rate Response During Exercise in Normals: Relevance for Rate Variable PacingPacing and Clinical Electrophysiology, 1993
- Standardized Informal Exercise Testing for Programming Rate Adaptive PacemakersPacing and Clinical Electrophysiology, 1991
- An office-based exercise protocol for predicting chronotropic response of activity-triggered, rate-variable pacemakersThe American Journal of Cardiology, 1989
- Comparison of exercise performance of six rate-adaptive right ventricular cardiac pacemakersThe American Journal of Cardiology, 1989
- Activity‐Sensing, Rate‐Responsive Pacing: Improvement in Myocardial Performance with ExercisePacing and Clinical Electrophysiology, 1985
- Atrial Synchronized Ventricular Pacing: Contribution of the Chronotropic Response to Improved Exercise PerformancePacing and Clinical Electrophysiology, 1983
- HAEMODYNAMIC EFFECT OF ATRIAL TRIGGERED VERSUS FIXED RATE PACING AT REST AND DURING EXERCISE IN COMPLETE HEART BLOCKActa Medica Scandinavica, 1975