Outcome, observer reliability, and patient preferences if CTA, MRA, or Doppler ultrasound were used, individually or together, instead of digital subtraction angiography before carotid endarterectomy
Open Access
- 1 July 2002
- journal article
- research article
- Published by BMJ in Journal of Neurology, Neurosurgery & Psychiatry
- Vol. 73 (1) , 21-28
- https://doi.org/10.1136/jnnp.73.1.21
Abstract
Objectives: To evaluate the accuracy of routinely available non-invasive tests (spiral computed tomographic angiography (CTA), time of flight magnetic resonance angiography (MRA), and colour Doppler ultrasound (DUS)), individually and together, compared with intra-arterial digital subtraction angiography (DSA) in patients with symptomatic tight carotid stenosis; and to assess the effect of substituting non-invasive tests for DSA on outcome, interobserver variability, and patient preference. Methods: Patients referred from a neurovascular clinic were subjected prospectively to DUS imaging. The operator was blind to symptoms. Patients with a tight carotid stenosis on the symptomatic side were admitted for DSA. CTA and MRA were performed during the admission. The CTA, MRA, and DSA films were each read independently by two of six experienced radiologists, blind to all other data. Results: 67 patients were included (34 had all four imaging procedures). DUS, CTA, and MRA all agreed with DSA in the diagnosis of operable v non-operable disease in about 80% of patients. CTA tended to underestimate (sensitivity 0.65, specificity 1.0), MRA to overestimate (sensitivity 1.0, specificity 0.57), and DUS to agree most closely with (sensitivity 0.85, specificity 0.71) the degree of stenosis as shown by DSA. When using any two of the three non-invasive tests in combination, adding the third if the first two disagreed would result in very few misdiagnoses (about 6%). MRA had similar interobserver variability to CTA (both worse than DSA). Patients preferred CTA over MRA and DSA. Conclusions: DUS, CTA, and MRA all show similar accuracy in the diagnosis of symptomatic carotid stenosis. No technique on its own is accurate enough to replace DSA. Two non-invasive techniques in combination, and adding a third if the first two disagree, appears more accurate, but may still result in diagnostic errors.Keywords
This publication has 13 references indexed in Scilit:
- Variability in Doppler ultrasound influences referral of patients for carotid surgeryEuropean Journal of Ultrasound, 2000
- Critical Appraisal of the Design and Reporting of Studies of Imaging and Measurement of Carotid StenosisStroke, 2000
- Contrast-enhanced MR angiography of the carotid bifurcationJournal of Magnetic Resonance Imaging, 1999
- Observer variation in the interpretation of intra-arterial angiograms and the risk of inappropriate decisions about carotid endarterectomy.Journal of Neurology, Neurosurgery & Psychiatry, 1996
- Spiral CT angiography and selective digital subtraction angiography of internal carotid artery stenosis.1996
- Evaluation of the quality of clinical research studies of magnetic resonance angiography: 1991‐1994Journal of Magnetic Resonance Imaging, 1996
- Noninvasive Carotid Artery Testing: A Meta-analytic ReviewAnnals of Internal Medicine, 1995
- Equivalence of measurements of carotid stenosis. A comparison of three methods on 1001 angiograms. European Carotid Surgery Trialists' Collaborative Group.Stroke, 1994
- Measurement of internal carotid artery stenosis from source MR angiograms.Radiology, 1994
- Complications of cerebral angiography in patients with symptomatic carotid territory ischaemia screened by carotid ultrasound.Journal of Neurology, Neurosurgery & Psychiatry, 1993